Background: Thrombin generation assays (TGAs) performed with calibrated automated thrombography (CAT) in the presence of thrombomodulin (TM) indicate plasma hypercoagulability in cirrhosis.
Objective: To evaluate, in the presence of TM, the new ST-Genesia automated device developed for improving TGA vs the previously used CAT method, with plasma samples of patients with cirrhosis.
Patients/methods: Platelet-poor plasma samples were prepared from citrated blood samples of 52 healthy controls and 85 patients with cirrhosis (severity evaluated using the Child-Pugh score [CP]). TGAs were performed using CAT with PPP-Reagent and ST-Genesia with the STG-ThromboScreen reagent, in the presence of TM. Endogenous thrombin potential (ETP) was chosen as the main parameter.
Results: Whatever the method, ETP values were higher in patients than in healthy controls. All patients identified as hypercoagulable with ST-Genesia and STG-ThromboScreen were found hypercoagulable with CAT and PPP-Reagent. Conversely, eight and ten patients in the CP-A and CP-B classes respectively were identified as hypercoagulable only with CAT. The use of ST-Genesia with the STG-ThromboScreen reagent with TM led to a bias, with higher ETP values for healthy controls and lower for patients compared with CAT. Crossover analysis (CAT with the STG-ThromboScreen reagent) evidenced a substantial effect of the STG-ThromboScreen reagent; the analyzer (including calibration and data analysis) plays a lesser role.
Conclusion: ST-Genesia evidences hypercoagulability in patients with cirrhosis when TG is studied in the presence of TM, but the results are not interchangeable with those obtained with CAT.
Keywords: blood coagulation; blood coagulation test; clinical laboratory techniques; liver cirrhosis; thrombomodulin.
© 2020 International Society on Thrombosis and Haemostasis.