A Sensor-Based Decision Support System for Transfemoral Socket Rectification

Sensors (Basel). 2021 May 28;21(11):3743. doi: 10.3390/s21113743.

Abstract

A decision support system (DSS) was developed that outputs suggestions for socket-rectification actions to the prosthetist, aiming at improving the fitness of transfemoral prosthetic socket design and reducing the time needed for the final socket design. For this purpose, the DSS employs a fuzzy-logic inference engine (IE) which combines a set of rectification rules with pressure measurements generated by sensors embedded in the socket, for deciding the rectification actions. The latter is then processed by an algorithm that receives, manipulates and modifies a 3D digital socket model as a triangle mesh formatted inside an STL file. The DSS results were validated and tested in an FEA simulation environment, by simulating and comparing the donning process among a good-fitting socket, a loose socket (poor-fit) and several rectified sockets produced by the proposed DSS. The simulation results indicate that volume reduction improves the pressure distribution over the stump. However, as the intensity of socket rectification increases, i.e., as volume reduction increases, high pressures appear in other parts of the socket which generate discomfort. Therefore, a trade-off is required between the amount of rectification and the balance of the pressure distributions experienced at the stump.

Keywords: FEA; fuzzy-logic inference engine; pressure sensors; prosthesis; socket rectification; transfemoral.

MeSH terms

  • Amputation Stumps
  • Artificial Limbs*
  • Computer Simulation
  • Prosthesis Design
  • Prosthesis Fitting

Grants and funding