Statistical Evaluation of Radiofrequency Exposure during Magnetic Resonant Imaging: Application of Whole-Body Individual Human Model and Body Motion in the Coil

Int J Environ Res Public Health. 2019 Mar 25;16(6):1069. doi: 10.3390/ijerph16061069.

Abstract

The accurate estimation of patient's exposure to the radiofrequency (RF) electromagnetic field of magnetic resonance imaging (MRI) significantly depends on a precise individual anatomical model. In the study, we investigated the applicability of an efficient whole-body individual modelling method for the assessment of MRI RF exposure. The individual modelling method included a deformable human model and tissue simplification techniques. Besides its remarkable efficiency, this approach utilized only a low specific absorption rate (SAR) sequence or even no MRI scan to generate the whole-body individual model. Therefore, it substantially reduced the risk of RF exposure. The dosimetric difference of the individual modelling method was evaluated using the manually segmented human models. In addition, stochastic dosimetry using a surrogate model by polynomial chaos presented SAR variability due to body misalignment and tilt in the coil, which were frequently occurred in the practical scan. In conclusion, the dosimetric equivalence of the individual models was validated by both deterministic and stochastic dosimetry. The proposed individual modelling method allowed the physicians to quantify the patient-specific SAR while the statistical results enabled them to comprehensively weigh over the exposure risk and get the benefit of imaging enhancement by using the high-intensity scanners or the high-SAR sequences.

Keywords: deformable human model; numerical simulation; specific absorption rate; stochastic dosimetry; transmit coil.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Electromagnetic Fields*
  • Female
  • Humans
  • Magnetic Resonance Imaging*
  • Male
  • Models, Theoretical
  • Radiation Exposure / analysis*
  • Radio Waves
  • Radiometry / methods*