Abstract

The stability of radiance measurements taken by the Sky Quality Meter (SQM)was tested under rapidly changing temperature conditions during exposure to a stable light field in the laboratory. The reported radiance was found to be negatively correlated with temperature, but remained within 7% of the initial reported radiance over a temperature range of -15 °C to 35 °C, and during temperature changes of -33 °C/h and +70 °C/h.This is smaller than the manufacturer's quoted unit-to-unit systematic uncertainty of 10%,indicating that the temperature compensation of the SQM is adequate under expected outdoor operating conditions.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Artifacts*
  • Atmosphere / analysis*
  • Environmental Monitoring / instrumentation*
  • Equipment Design
  • Equipment Failure Analysis
  • Photometry / instrumentation*
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Temperature
  • Transducers*