FluoroTome 1: An Apparatus for Tomographic Imaging of Radio-Fluorogenic (RFG) Gels

Polymers (Basel). 2019 Oct 23;11(11):1729. doi: 10.3390/polym11111729.

Abstract

Radio-fluorogenic (RFG) gels become permanently fluorescent when exposed to high-energy radiation with the intensity of the emission proportional to the local dose of radiation absorbed. An apparatus is described, FluoroTome 1, that is capable of taking a series of tomographic images (thin slices) of the fluorescence of such an irradiated RFG gel on-site and within minutes of radiation exposure. These images can then be compiled to construct a 3D movie of the dose distribution within the gel. The historical development via a laboratory-bench prototype to a readily transportable, user-friendly apparatus is described. Instrumental details and performance tests are presented.

Keywords: 3D radiation dosimetry; fluorescent gels; polymer-gel radiation dosimetry; radio-fluorogenic (RFG) gel; tomographic fluorescence imaging.