A software for quantification of vessel density in glaucoma: An OCT-Angiography study

J Fr Ophtalmol. 2021 Mar;44(3):376-381. doi: 10.1016/j.jfo.2020.06.038. Epub 2021 Jan 26.

Abstract

Purpose: To assess the capillary vessel density in the peripapillary region of glaucoma patients in comparison to controls using automated software.

Methods: We performed an observational, cross-sectional case-control study with 72 eyes of glaucoma patients (with strict selection criteria) and 30 age-matched healthy controls. We used an optical coherence tomography angiography device (Triton, Topcon) with scans of 4.5×4.5mm centered on the optic nerve head, excluding images of poor quality. Since this device does not quantify vessel density, we wrote software (with Visual Studio©) that allowed us to remove large blood vessels and quantify the capillary density in the peripapillary region. We tested the ability of OCT-A to differentiate glaucoma from controls using the receiver operating characteristic curve (ROC). We also verified whether VD was correlated with visual field mean deviation.

Results: VD was significantly lower in glaucoma than in controls, notably the whole-image VD of 72.45±7.45% in glaucoma and 77.87±3.77% in controls (P<0.001). VD values were correlated with visual fields (Spearman correlation of 0.33; P 0.0017). ROC was moderate (color VD of 0.629±0.143), but better than that of the visual field (0.229±0.077) or the retinal nerve fiber layer thickness assessed by OCT (0.326±0.064).

Conclusions: This software is useful for quantifying vessel density at the optic disc, peripapillary and capillary levels and for documenting glaucomatous changes in vascularization. OCT-A might be useful in diagnosis of glaucoma.

Keywords: Case-control; Contrôle des cas; Densité des vaisseaux; Diagnosis; Diagnostic; Glaucoma; Glaucome; Logiciel; OCT-Angiographie; OCT-Angiography; OCTA; Software; Vessel density.

Publication types

  • Observational Study

MeSH terms

  • Case-Control Studies
  • Cross-Sectional Studies
  • Fluorescein Angiography
  • Glaucoma* / diagnostic imaging
  • Glaucoma, Open-Angle*
  • Humans
  • Intraocular Pressure
  • Optic Disk* / diagnostic imaging
  • Retinal Vessels / diagnostic imaging
  • Software
  • Tomography, Optical Coherence