Thermal Characterization of Dynamic Silicon Cantilever Array Sensors by Digital Holographic Microscopy

Sensors (Basel). 2017 May 23;17(6):1191. doi: 10.3390/s17061191.

Abstract

In this paper, we apply a digital holographic microscope (DHM) in conjunction with stroboscopic acquisition synchronization. Here, the temperature-dependent decrease of the first resonance frequency (S₁(T)) and Young's elastic modulus (E₁(T)) of silicon micromechanical cantilever sensors (MCSs) are measured. To perform these measurements, the MCSs are uniformly heated from T₀ = 298 K to T = 450 K while being externally actuated with a piezo-actuator in a certain frequency range close to their first resonance frequencies. At each temperature, the DHM records the time-sequence of the 3D topographies for the given frequency range. Such holographic data allow for the extracting of the out-of-plane vibrations at any relevant area of the MCSs. Next, the Bode and Nyquist diagrams are used to determine the resonant frequencies with a precision of 0.1 Hz. Our results show that the decrease of resonance frequency is a direct consequence of the reduction of the silicon elastic modulus upon heating. The measured temperature dependence of the Young's modulus is in very good accordance with the previously-reported values, validating the reliability and applicability of this method for micromechanical sensing applications.

Keywords: digital holography; micromechanical cantilever sensors; temperature coefficient of elastic modulus; temperature coefficient of resonance frequency; thermal load.