Interobserver Agreement and Plane-Dependent Intraobserver Variability of Shear Wave Sonoelastography in the Differential Diagnosis of Ectopic Thymus Tissue

J Clin Med. 2021 Jan 9;10(2):214. doi: 10.3390/jcm10020214.

Abstract

Shear wave elastography (SWE) has been demonstrated to be a useful tool in the differential diagnosis of ectopic thymus tissues (ETs), providing quantitative values of the shear wave stiffness (SWS) of both ETs and adjacent thyroid tissue. However, no data are available on the potential influence of the imaging plane (transverse vs. longitudinal) on the obtained SWS and shear wave ratio (SWR) values in SWE of these tissues. Moreover, no reports on the interobserver repeatability of SWE were published in regard to ETs. The aim of this study has been to evaluate the potential influence of the examination plane-transverse vs. longitudinal-on the SWS and SWR results, as well as to determine whether SWE of ETs is subjected to interobserver variability. SWE was demonstrated to have high inter- and intraobserver agreement in the evaluation of ETs and adjacent thyroid tissue. Significant differences between SWS values, but not SWR values, obtained in the transverse and longitudinal planes were observed. This phenomenon is probably a result of anisotropy-related artifacts and does not reduce the reliability of the method. SWE operators should be aware of the presence of plane-dependent artifacts to properly interpret the obtained results.

Keywords: ectopic thymus; interobserver variability; shear wave elastography; thyroid; ultrasound.