Feasibility of Ultra-High Resolution Supra-Aortic CT Angiography: An Assessment of Diagnostic Image Quality and Radiation Dose

Tomography. 2021 Nov 1;7(4):711-720. doi: 10.3390/tomography7040059.

Abstract

(1) Background: To evaluate diagnostic image quality and radiation exposure of ultra-high resolution cerebral Computed-Tomography (CT) angiography (CTA) obtained on an ultra-high resolution computed tomography scanner (UHR-CT). (2) Methods: Fifty consecutive patients with UHR-CTA were enrolled. Image reconstruction was processed with a 1024 × 1024 matrix and a slice thickness of 0.25 mm. Quantitative analyses comprising CT values, contrast-noise ratio (CNR) and signal-to-noise ratio (SNR) were performed. Subjective assessment of image quality, vessel contrast, noise, artefacts and delineation of different sized vessels were assessed by two readers on a 4-point scale. Radiation exposure was determined. (3) Results: Hounsfield values (ACI: 461.8 ± 16.8 HU; MCA: 406.1 ± 24.2 HU; BA: 412.2 ± 22.3 HU), SNR (ACI: 35.4 ± 13.1; MCA: 20.8 ± 12.4; BA: 23.7 ± 12.9) and CNR (ACI: 48.7 ± 21; MCA: 63.9 ± 26.9; BA: 48.1 ± 21.4) were remarkably high in all segments. Subjective analysis by two raters (fair agreement, k = 0.26) indicated excellent image qualities (image quality = 4; contrast = 4; noise = 3; artefacts = 4).Our analysis revealed a notably high traceability of the cerebral perforators (3 Points). Radiation exposure was at moderate dose levels (effective dose = 2.5 ± 0.6mSv). (4) Conclusions: UHR-CTA generates highly valuable image qualities that allow the depiction of vessels including cerebral perforators at acceptable dose levels. The UHR-CTA may therefore enhance the detection of small cerebral pathologies and may improve interpretability, especially in settings where high image qualities are crucial for the diagnostic accuracy.

Keywords: cerebral arteries; computed tomography; computed tomography angiography; contrast media; image enhancement; radiation dosage; stroke.

MeSH terms

  • Cerebral Angiography / methods
  • Computed Tomography Angiography* / methods
  • Feasibility Studies
  • Humans
  • Radiation Dosage
  • Radiographic Image Interpretation, Computer-Assisted* / methods
  • Tomography, X-Ray Computed / methods