The aim of this study was to evaluate the measurement of C-peptide, insulin, sex hormone binding globulin (SHBG), and dehydroepiandrosterone sulfate (DHEAS), tests of specific clinical value that are not frequently measured in routine clinical laboratories, carried out using an automated system, AIA 2000 (Tosoh). The obtained data demonstrated that the evaluated system, characterized by fluorescence system detector provides satisfactory analytical performance comparable to those of different instrument that use the well known and widely diffused chemiluminescent principle (Immulite 2000). These characteristics allow an excellent comparability between methods for the measurement of same specific hormones not frequently used in clinical practice.