Modeling the relationship between psychophysical perception and electrically evoked compound action potential threshold in young cochlear implant recipients: clinical implications for implant fitting

Clin Neurophysiol. 2004 Dec;115(12):2811-24. doi: 10.1016/j.clinph.2004.06.024.

Abstract

Objective: In cochlear implant recipients, the threshold of the electrically evoked compound action potential (ECAP) has been shown to correlate with the perceptual detection threshold and maximum comfortable loudness levels (respectively, T- and C-levels) used for implant programming. Our general objective was to model the relationship between ECAP threshold and T/C-levels by taking into account their relative changes within each subject. In particular, we were interested in investigating further the validity of ECAP threshold as a predictor of psychophysical levels, depending on intra-cochlear electrode location and time of testing (from 1 to 18 months post-implantation).

Methods: A total of 370 ECAP thresholds, measured in 49 children, using a Nucleus 24 cochlear implant, were compared with the corresponding T- and C-levels obtained at the same visit, for the same electrode. Response profiles for the whole group of patients were modeled across four test electrodes spaced equally along the electrode array from base towards apex. A linear regression model was constructed and the quality of the ECAP threshold-based predictions was assessed by testing for correlation between measured and predicted psychophysics. Comparison was made with a more simplistic model (described here as the 'parallel profiles method') stipulating, within each subject, a 1 microA increase in psychophysical levels for every 1 microA increase in ECAP threshold.

Results: Offset between ECAP threshold and psychophysics profiles was found to vary significantly along the electrode array for the T-, but not for the C-level. In contrast with the parallel profiles method, our regression model predicted, within each subject, an average increase of 0.23 microA (95% confidence interval: 0.18-0.28) in T-level for every 1 microA increase in ECAP threshold. This correction improved the quality of T-level prediction when our model was run using measured T-level and ECAP threshold from a reference electrode (r=0.77 vs. r=0.62). The shorter the distance between the electrode for which T-level was predicted and the one used as reference, the stronger the correlation between measured and predicted T-levels. In addition, poorer T-level predictions were obtained at the basal end of the array during the first 3 months post-implantation. In contrast to T-level, individual changes in C-level with ECAP threshold exhibited heterogeneous patterns across subjects so that no common coefficient could account for these changes. However, applying the parallel profiles method led to high-quality C-level prediction.

Conclusions and significance: The results suggest that covariation between ECAP thresholds and psychophysics plays a decisive role in the relationship of ECAP threshold with T-, but not with C-level. Therefore, our regression model and the parallel profiles method should both be used for predicting, respectively, the T- and the C-levels. Although the predictability of our regression model seems to be better for middle and apical electrodes, its utilization should be extended to basal electrodes after 6 months' implant use.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Action Potentials
  • Auditory Perception*
  • Auditory Threshold
  • Child
  • Child, Preschool
  • Cochlear Implantation / methods*
  • Evoked Potentials, Auditory*
  • Female
  • Humans
  • Infant
  • Male
  • Models, Theoretical*
  • Psychophysics*
  • Regression Analysis