Aortic valve replacement with the Biocor PSB stentless xenograft

Ann Thorac Surg. 1998 Aug;66(2):425-30. doi: 10.1016/s0003-4975(98)00345-2.

Abstract

Background: The midterm clinical results after aortic valve replacement with the Biocor PSB stentless xenograft on all patients operated between October 1992 and October 1996 were reviewed.

Methods: One hundred six patients, aged 70+/-6 years, had aortic valve replacement for aortic stenosis (67%), regurgitation (11%), or both (22%). Associated procedures were done in 49 patients (46%), including coronary artery bypass in 30 patients, mitral valve repair/replacement in 16, and ascending aorta replacement in 5 patients. Aortic cross-clamp and cardiopulmonary bypass times were 96+/-24 and 129+/-31 minutes, respectively.

Results: There were 3 (3%) early deaths due to low output (2 patients) and cerebrovascular accident (1 patient). Follow-up of survivors ranged from 6 to 66 months (mean, 39+/-14 months). Survival was 94%+/-2% and 90%+/-3% at 1 and 5 years. There were 5 late deaths due to cardiac cause (2), cancer (2), and pulmonary embolism (1 patient). No patient had structural valve deterioration, whereas 100% and 95%+/-3% were free from valve-related events at 1 and 5 years. There were two reoperations due to narrowing of the left coronary ostium and endocarditis, with an actuarial freedom from reoperation of 99%+/-1% and 98+/-1% at 1 and 5 years, respectively. Functional results demonstrated a mean peak transprosthetic gradient of 16+/-12 mm Hg, with only 1 patient (1%) with a 55 mm Hg gradient. No cases of valve regurgitation greater than mild were recorded at follow-up. Assessment of New York Heart Association functional class demonstrated a significant improvement (2.9+/-0.6 versus 1.4+/-0.7; p=0.01). All patients were free from anticoagulation.

Conclusions: Aortic valve replacement using the Biocor PSB stentless xenograft offers excellent midterm survival, negligible valve deterioration, and a very low rate of valve-related events, which are comparable to estimates reported with other models of stentless xenografts and currently available stented xenografts. Hemodynamic performance is favorable and quality of life satisfactory.

MeSH terms

  • Aged
  • Aortic Valve / surgery*
  • Aortic Valve Insufficiency / surgery
  • Aortic Valve Stenosis / surgery
  • Bioprosthesis*
  • Cardiopulmonary Bypass
  • Coronary Artery Bypass
  • Endocarditis, Bacterial / etiology
  • Heart Valve Prosthesis Implantation*
  • Hemodynamics
  • Humans
  • Mitral Valve / surgery
  • Postoperative Complications
  • Quality of Life
  • Reoperation
  • Survival Rate