The use of improved radiochromic film for in vivo quality assurance of high dose rate brachytherapy

Med Phys. 1998 Jul;25(7 Pt 1):1217-21. doi: 10.1118/1.598311.

Abstract

GAFChromic film has become increasingly popular for radiation dosimetry. In this study we explore the use of GAFChromic film as an in vivo dosimeter for quality assurance (QA) of fractionated high dose rate 192Ir treatments. Accuracy of dose distribution is explored for the simple vaginal cylinder geometry for which the dose can be easily calculated for comparison. Source dwell times for several patients were optimized to deliver 500 cGy at 0.5 cm from the surface of the vaginal cylinder applicator using a commercial treatment planning system. GAFChromic film was taped to the vaginal cylinder applicator and was enclosed in a leak proof rubber sleeve prior to its insertion. Optical densities were measured along the film at 2 mm spacing, using a densitometer with filtered red light. Density corrections for transient film darkening effects were made and optical densities were converted to absorbed dose in cGy. In vivo patient dose distribution measured for different patients and different fractions were compared with the calculated values along the applicator surface. The variation between the calculated and measured dose was +/- 10%. the reproducibility of dose measurement for different fraction was within +/- 5%. This study demonstrates the potential usefulness of the film as an in vivo for brachytherapy QA.

Publication types

  • Comparative Study

MeSH terms

  • Administration, Intravaginal
  • Brachytherapy / instrumentation*
  • Brachytherapy / methods
  • Brachytherapy / standards*
  • Calibration
  • Densitometry
  • Female
  • Humans
  • Iridium Radioisotopes / therapeutic use
  • Models, Theoretical
  • Phantoms, Imaging
  • Quality Control
  • Radiometry / instrumentation*
  • Radiometry / methods
  • Radiotherapy Planning, Computer-Assisted / statistics & numerical data
  • Reproducibility of Results
  • X-Ray Film*

Substances

  • Iridium Radioisotopes