Femoral nerve lesion in total hip replacement: an experimental study

Arch Orthop Trauma Surg. 1998;117(3):153-5. doi: 10.1007/s004020050216.

Abstract

A total of 20 hip joints of 10 non-fixed corpses were examined within 48 h of death to measure the pressure below the inguinal ligament simulating the surgical conditions during total hip arthroplasty. The purpose of this study was to assess the influence of various leg positions and insertion techniques of retractors during the surgical procedure for total hip replacement in order to detect supposed causes for indirect pressure injuries of the femoral nerve. The obtained results verified no increase of pressure in the inguinal canal which could explain an indirect injury of the femoral nerve. If the retractor is inserted correctly at the anterior acetabular rim, the pressure in the lacuna musculorum can even be reduced, and furthermore, the femoral nerve is protected by the iliopsoas muscle. Femoral nerve lesions which have been published so far can only be explained by an incorrect use of instruments or implants (e.g., screws, cement, acetabular cup) or an extreme postoperative leg length discrepancy.

MeSH terms

  • Arthroplasty, Replacement, Hip / adverse effects*
  • Arthroplasty, Replacement, Hip / instrumentation
  • Arthroplasty, Replacement, Hip / methods
  • Cadaver
  • Femoral Nerve / injuries*
  • Humans
  • Monitoring, Intraoperative / methods
  • Peripheral Nervous System Diseases / etiology
  • Peripheral Nervous System Diseases / prevention & control
  • Posture
  • Pressure
  • Sensitivity and Specificity
  • Wounds, Nonpenetrating / etiology*
  • Wounds, Nonpenetrating / prevention & control