Alanine dosimetry of proton therapy beams

Med Phys. 1997 Mar;24(3):447-53. doi: 10.1118/1.597911.

Abstract

An analysis of some of the properties of the ESR-alanine dosimetry in therapeutic proton beams is reported. Thin alanine-based detectors (1 and 2 mm thick pellets and 0.25 mm thick films) were tested in a clinical 62 MeV proton beam. The response of the alanine detectors in unmodulated and modulated proton beams was studied in tissue equivalent phantoms. The dose assessed by alanine was compared to the dose provided by a Markus parallel plate ionization chamber, used for reference dosimetry. Experiments in the 5-250 Gy dose range showed linearity of alanine dose response and no significant differences in the alanine response per unit dose to gammas and protons. Agreement within the experimental uncertainties was found between the alanine and the Markus chamber depth dose curves, including the Bragg peak region.