Compact tunable Cr:LiSAF laser for infrared matrix-assisted laser desorption/ionization

Rapid Commun Mass Spectrom. 1997;11(4):393-7. doi: 10.1002/(SICI)1097-0231(19970228)11:4<393::AID-RCM868>3.0.CO;2-S.

Abstract

A tunable Cr:LiSAF laser-pumped optical parametric oscillator was used for mid-infrared matrix-assisted laser desorption/ionization (MALDI) mass spectrometry experiments. The mass spectra of substance P, bovine insulin and pd(T)10 in the 2.88-2.96 microns range showed excellent single shot signal quality (signal-to-noise-ratio, resolution) but poor shot-to-shot reproducibility. The reproducibility is expected to improve with more stable laser design. No correlation was found between the absorption spectrum of the matrix and the MALDI response.

Publication types

  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Animals
  • Cattle
  • Infrared Rays
  • Insulin / analysis
  • Lasers*
  • Nucleic Acids / analysis
  • Reproducibility of Results
  • Spectrometry, Mass, Matrix-Assisted Laser Desorption-Ionization / instrumentation*
  • Spectrophotometry, Infrared
  • Substance P / analysis