Cancellation of metal-induced MRI artifacts with dual-component paramagnetic and diamagnetic material: mathematical modelization and experimental verification

J Magn Reson Imaging. 1996 Nov-Dec;6(6):936-8. doi: 10.1002/jmri.1880060615.

Abstract

A mathematical model of dual-component paramagnetic and diamagnetic material to cancel metal-induced MRI artifacts was developed and verified experimentally. The magnetization produced by metallic material and then the gradient linearity distortion can be cancelled by using such materials with opposing paramagnetic and diamagnetic properties. This concept of dual-component materials provides a novel solution to the problem of MRI artifacts.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Artifacts*
  • Computer Simulation*
  • Magnetic Resonance Imaging / instrumentation
  • Magnetic Resonance Imaging / methods*
  • Magnetic Resonance Imaging / statistics & numerical data
  • Magnetics*
  • Metals*
  • Prostheses and Implants*

Substances

  • Metals