Calibration of the Ektachem Amylase method using human reference materials and the Phadebas Blue Starch method as an interim reference method

Scand J Clin Lab Invest. 1996 Oct;56(6):505-9. doi: 10.3109/00365519609088806.

Abstract

In a multilaboratory study a consensus was established to calibrate the Ektachem Amylase method to reproduce the results of the Phadebas Blue Starch method. The calibration graph has a slope = 3.39 and intercept = 25. For the Ektachem Amylase method the reaction-rate ratio between salivary and pancreatic amylase was calculated to be 0.89, relative to that of the Phadebas Blue Starch method. A calibration value for the Nordic Amylase Calibrator to be used on Ektachem analysers was determined to be close to 469 U l-1 for the current batch. However, since the difference from the stated value of 460 U l-1 is negligible, the authors recommend the use of the stated value for this and future batches of the Nordic Amylase Calibrator. An error of around 10% introduced by the presence of salivary amylase is comparable to the methods accepted by the Scandinavian Committee on Enzymes. Introduction of a consensus calibration reduced the interlaboratory variation by up to 40% at all levels.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Amylases / analysis*
  • Biomarkers / analysis
  • Evaluation Studies as Topic
  • Humans
  • Pathology, Clinical / methods*
  • Reference Standards
  • Reference Values

Substances

  • Biomarkers
  • Amylases