Comparison of imaging accuracy at different MRI units based on phantom measurements

Radiother Oncol. 1995 Dec;37(3):221-4. doi: 10.1016/0167-8140(95)01657-0.

Abstract

In this investigation standardized on-site phantom measurements at 27 MRI units were carried out. A scoring system was developed for rating both symmetry and distortion of the image. This study shows, with regard to the clinical use of MRI-assisted treatment planning in radiotherapy, that in more than 85% of all planes (transversal, sagittal, coronal) the MRI units were equivalent to the standard known from literature.

Publication types

  • Comparative Study

MeSH terms

  • Artifacts
  • Humans
  • Magnetic Resonance Imaging / instrumentation*
  • Models, Anatomic*
  • Quality Control
  • Radiotherapy Planning, Computer-Assisted / instrumentation*
  • Reference Values