Fluid dynamics of venous valve closure

Ann Biomed Eng. 1995 Nov-Dec;23(6):750-9. doi: 10.1007/BF02584474.

Abstract

In vitro experiment was performed on a stented bovine jugular vein valve (VV, 14 mm I.D. x 2 cm long) and a stentless bovine jugular vein valve conduit (10 mm I.D. x 6 cm long) in a hydraulic flow loop with a downstream oscillatory pressure source to mimic respiratory changes. Simultaneous measurements were made on the valve opening area, conduit and sinus diameter changes using a specially designed laser optic system. Visualization of flow fields both proximal and distal to the venous valve, and the valve opening area were simultaneously recorded by using two video cameras. Laser Doppler anemometer surveys were made at three cross sections: the valve inlet, the valve exist, and 2 cm downstream of the venous valve to quantity flow reflux at valve closure. The experiment confirmed that the VV is a pressure-operated rather than a flow-driven device and that little or no reflux is needed to close the valve completely. The experiment further demonstrated that the VV sinus expands rapidly against back pressure, a critical character to consider in venous prosthesis design.