Autoradiography for iodine-125 seeds

Int J Radiat Oncol Biol Phys. 1993 Apr 2;25(5):881-4. doi: 10.1016/0360-3016(93)90319-q.

Abstract

To study the interior design of model 6702 and 6711 iodine-125 seeds contact autoradiographs were performed using mammography film. Improved resolution was obtained using a pin-hole camera with a hole of 0.1 mm x 0.1 mm. With these techniques, qualitative determination of the relative activity distribution within each seed was possible. The number of the activated resin spheres and the positions of the centers of these spheres can be exactly determined. A model calculation shows, that variations in the arrangement of the activated spheres within a seed have a moderate influence on the dose distribution at source distances below 10 mm. Knowing the exact source configuration may be useful when comparing dose calculations with measured data for model 6702 125I seeds which are currently employed in ophthalmic plaque and implant therapy of other tumors.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Autoradiography
  • Female
  • Humans
  • Iodine Radioisotopes*
  • Mammography
  • Radiation Dosage