The accuracy of six impression techniques for osseointegrated implants

J Prosthodont. 1993 Jun;2(2):75-82. doi: 10.1111/j.1532-849x.1993.tb00387.x.

Abstract

Purpose: Six impression techniques were evaluated using tapered and square impression copings.

Materials and methods: The absolute distortion was measured using a cast metal impression tray with fiduciary reference points external to the impression material. Measurements of the x, y, z coordinates on the master cast and the impressions were directly made with a travelling digitizing microscope. The difference between the coordinates of each of six sites in the impression and the corresponding reference site were compared.

Results: There was no significant difference between the techniques for the square copings but that there was a significant loss of accuracy in the z-axis with the tapered copings.

Conclusions: No significant differences were noted.

Publication types

  • Comparative Study

MeSH terms

  • Analysis of Variance
  • Dental Abutments
  • Dental Implants*
  • Dental Impression Materials
  • Dental Impression Technique*
  • Humans
  • Models, Dental
  • Reproducibility of Results