The quality of photographs produced by the non-mydriatic fundus camera in a screening programme for diabetic retinopathy: a 1 year prospective study

Eye (Lond). 1993:7 ( Pt 6):787-90. doi: 10.1038/eye.1993.185.

Abstract

A prospective study was performed on the quality of photographs produced by the non-mydriatic fundus camera used in a hospital-based screening programme for diabetic retinopathy. In 1 year 981 binocular patients were photographed. A photograph of acceptable quality was obtained from 90.5% of eyes and 84.4% of patients had an acceptable photograph of both eyes. The photograph of the second eye was more often unacceptable than that of the first. This tendency was significant in females (p = 0.0196) and when considering the sexes together (p = 0.0044), but not significant in males (p = 0.1042). Photographs of unacceptable quality were obtained significantly more often in patients aged over 55 years for both right and left eyes (p = 0.001). An overall improvement in photographic quality might be achieved by allowing full recovery of flash-induced pupil constriction before taking the second photograph and by dilating those aged over 55 years.

MeSH terms

  • Adult
  • Aged
  • Diabetic Retinopathy / pathology*
  • Female
  • Fundus Oculi
  • Humans
  • Light
  • Male
  • Middle Aged
  • Photography*
  • Prospective Studies
  • Pupil / physiology
  • Retina / pathology
  • Sex Factors