Criteria for anatomical compatibility of the total artificial heart: computerized three-dimensional modeling of the cardiovascular anatomy

Artif Organs. 1993 Dec;17(12):1022-35. doi: 10.1111/j.1525-1594.1993.tb03185.x.

Abstract

A quantitative study of cardiovascular anatomy was performed by obtaining three-dimensional reconstructions from regular computed tomography scan images in 15 patients, all candidates for heart transplantation. Volumetric estimates of the cardiovascular structures were obtained from these three-dimensional reconstructions using data directly related to total artificial heart (TAH) implantations. By using computerized three-dimensional modeling of these structures, reproducible measurements of the parameters defining the shape and the anatomical connections of the intrathoracic space available for TAH implantation could be derived. The results are intended to be used for both technical and clinical applications such as computer-assisted drawing of the pericardial cavity and the anatomical connections (useful for improving the design of TAH) and combined statistical calculations (multiple regressions, cluster algorithm) of the measurement results, which will then enable the best selection to be made among two or three TAH models for each patient.