Ultrasound velocities for axial eye length measurement

J Cataract Refract Surg. 1994 Sep;20(5):554-62. doi: 10.1016/s0886-3350(13)80237-4.

Abstract

Since 1974, I have used individual sound velocities for each eye condition encountered for axial length measurement. The calculation results in 1,555 M/sec for the average phakic eye. A slower speed of 1,549 M/sec was found for an extremely long (30 mm) eye and a higher speed of 1,561 M/sec was noted for an extremely short (20 mm) eye. This inversely proportional velocity change can best be adjusted for by measuring the phakic eye at 1,532 M/sec and correcting the result by dividing the square of the measured axial length (AL1,532)2 by the difference of the measured axial length (AL1,532) minus 0.35 mm. A velocity of 1,534 M/sec was found for all aphakic eyes regardless of their length, and correction is clinically significant. The velocity of an eye containing a poly(methyl methacrylate) intraocular lens is not different from an average phakic eye but it does magnify the effect of axial length change. I recommend measuring the pseudophakic eye at 1,532 M/sec and adding to the result (AL1,532), + 0.04 + 44% of the IOL thickness. The speed for an eye with a silicone IOL was found to be 1,476 M/sec (or AL1,532 + 0.04 - 56% of IOL thickness) and for glass, 1,549 M/sec (or AL1,532 + 0.04 + 75% of IOL thickness). A speed of 1,139 M/sec was found for a phakic eye with silicone oil filling most of the vitreous cavity and 1,052 M/sec for an aphakic eye filled with oil. For varying volumes of oil, each eye should be calculated individually. The speed was 534 M/sec for phakic eyes filled with gas. Eyes containing a silicone IOL or oil or gas will create clinically significant errors (3 to 10 diopters) if the sound velocity is not corrected.

MeSH terms

  • Aphakia, Postcataract / diagnostic imaging
  • Cornea / anatomy & histology
  • Cornea / diagnostic imaging
  • Eye / anatomy & histology*
  • Eye / diagnostic imaging*
  • Glass
  • Humans
  • Lenses, Intraocular
  • Mathematics
  • Methylmethacrylate
  • Methylmethacrylates
  • Optics and Photonics
  • Silicone Elastomers
  • Silicone Oils
  • Sulfur Hexafluoride
  • Ultrasonography / methods

Substances

  • Methylmethacrylates
  • Silicone Elastomers
  • Silicone Oils
  • Methylmethacrylate
  • Sulfur Hexafluoride