[The growth behavior of mouse fibroblasts on intraocular lens surface of various silicone and PMMA materials]

Ophthalmologe. 1994 Aug;91(4):521-5.
[Article in German]

Abstract

Experience with intraocular lenses (IOL) made of PMMA dates back ca. 40 years, while silicone IOLs have been in use for only about 10 years. The biocompatibility of PMMA and silicone caoutchouc was tested in a comparative study investigating the growth of mouse fibroblasts on different IOL materials. Spectrophotometric determination of protein synthesis and liquid scintillation counting of DNA synthesis were carried out. The spreading of cells was planimetrically determined, and the DNA synthesis of individual cells in direct contact with the test sample was tested. The results showed that the biocompatibility of silicone lenses made of purified caoutchouc is comparable with that of PMMA lenses; there is no statistically significant difference. However, impurities arising during material synthesis result in a statistically significant inhibition of cell growth on the IOL surfaces.

Publication types

  • Comparative Study
  • English Abstract

MeSH terms

  • Animals
  • Cell Division / physiology*
  • Cell Line
  • DNA Replication / physiology
  • Fibroblasts / cytology
  • Lenses, Intraocular*
  • Materials Testing
  • Methylmethacrylates*
  • Mice
  • Prosthesis Design
  • Silicone Elastomers*
  • Surface Properties

Substances

  • Methylmethacrylates
  • Silicone Elastomers