Fluid dynamic study of the moving actuator type total artificial heart by image-processing technique

Med Biol Eng Comput. 1994 May;32(3):253-7. doi: 10.1007/BF02512519.

Abstract

Flow velocity inside the artificial ventricle of a moving actuator type total artificial heart is investigated by the image-processing technique. Optical flow is estimated from two sequential images of flow patterns to find flow-velocity distribution. The region with high velocity inside the artificial ventricle is found from the estimated flow-velocity distribution. Fluid velocity in a simple glass conduit is analysed and can be estimated by calculating optical flow.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Blood Flow Velocity
  • Heart, Artificial*
  • Hemorheology*
  • Humans
  • Image Processing, Computer-Assisted*
  • Models, Cardiovascular
  • Ventricular Function