Determination of ammonium and L-glutamine in hybridoma cell cultures by sequential flow injection analysis

Cytotechnology. 1994;14(3):177-82. doi: 10.1007/BF00749614.

Abstract

A flow injection analytical system based on a gas diffusion membrane module for ammonia and an ammonium flow-through potentiometric detector has been set up for measurement of L-glutamine and ammonium ions in hybridoma cell cultures. The main feature of the system is that the same basic analytical concept and equipment is used in both measurements, the only difference being for the determination of L-glutamine, in which the sample flows through an immobilized glutaminase cartridge. The conditions to enable the performance of both analysis consecutively, avoiding potential interferences by unwanted deamination of other compounds in the samples, have been determined. Finally, the proposed system has been compared with reference analytical methods for batch hybridoma cell culture experiments.

MeSH terms

  • Animals
  • Diffusion
  • Flow Injection Analysis / methods*
  • Glutamine / analysis*
  • Hybridomas / chemistry*
  • Mice
  • Potentiometry
  • Quaternary Ammonium Compounds / analysis*
  • Reproducibility of Results

Substances

  • Quaternary Ammonium Compounds
  • Glutamine