Monitoring of blood-10B concentration for boron neutron capture therapy using prompt gamma-ray analysis

Acta Oncol. 1995;34(4):517-23. doi: 10.3109/02841869509094017.

Abstract

The aim of the present study was to monitor the blood-10B concentration of laboratory dogs receiving boron neutron capture therapy, in order to obtain optimal agreement between prescribed and actual dose. A prompt gamma-ray analysis system was developed for this purpose at the High Flux Reactor in Petten. The technique was compared with inductively coupled plasma-atomic emission spectrometry and showed good agreement. A substantial variation in 10B clearance pattern after administration of borocaptate sodium was found between the different dogs. Consequently, the irradiation commencement was adjusted to the individually determined boron elimination curve. Mean blood-10B concentrations during irradiation of 25.8 +/- 2.2 micrograms/g (1 SD, n = 18) and 49.3 +/- 5.3 micrograms/g (1 SD, n = 17) were obtained for intended concentrations of 25 micrograms/g and 50 micrograms/g, respectively. These variations are a factor of two smaller than irradiations performed at a uniform post-infusion irradiation starting time. Such a careful blood-10B monitoring procedure is a prerequisite for accurately obtaining such steep dose-response curves as observed during the dog study.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Boron / blood*
  • Boron Neutron Capture Therapy*
  • Dogs
  • Dose-Response Relationship, Radiation
  • Gamma Rays*
  • Half-Life
  • Isotopes
  • Neutron Activation Analysis / methods
  • Radiotherapy Dosage

Substances

  • Isotopes
  • Boron