Myopic keratomileusis by excimer laser on a lathe

J Refract Corneal Surg. 1994 Sep-Oct;10(5):575-81.

Abstract

Background: We designed an excimer laser keratomileusis delivery system to increase the regularity of the refractive cut surface and allow greater precision in the level and shape of the ablated zone.

Methods: A parallel faced corneal disc was produced by microkeratectomy from six human eyes and surgical keratectomy in 12 beagle corneas. A 193-nanometer excimer laser that was used to project an oval beam onto the corneal disc was rotated on a flat surface to ensure overlapping of the ovally ablated areas between pulses.

Results: Electron microscopy of eye bank lenticules demonstrated a circular smooth regularly concave ablation zone. Histological examination of nine clear corneas confirmed thinning of the stroma without fibroblastic reaction and no epithelial hypertrophy. Mean preoperative corneal power of 43.15 +/- 2.18 decreased postoperatively to 33.61 +/- 2.34.

Conclusions: The new technique of excimer laser keratomileusis has the advantage of a cut surface smoother and the clear zone is devoid of the stepwise concavity and irregularity seen in diaphragm based photoablation delivery systems.