[Radiation exposure in diagnostic CT: comparison between spiral CT and standard CT]

Aktuelle Radiol. 1995 Sep;5(5):289-92.
[Article in German]

Abstract

CT-Dose-Index of Standard-CT corresponds to CT-Dose-Effect-Index of Spiral-CT. Spiral-CT provides a continuous data acquisition with well balanced dose profiles. Increasing pitch reduces radiation dose and nevertheless an unbroken reconstruction of scanned volume is possible. In contrast to Standard-CT, small parts at both ends of the volume scan are not reconstructed. Volume-Dose-Index therefore is elevated in short scan times but corresponds to that of Standard-CT in long scan times. 1 mm slices cause more radiation exposure due to their broadened dose profiles.

Publication types

  • Comparative Study
  • English Abstract

MeSH terms

  • Equipment Design
  • Humans
  • Radiation Dosage
  • Radiometry / instrumentation*
  • Tomography, X-Ray Computed / instrumentation*