Feasibility of a novel augmented reality overlay for cervical screw placement in phantom spine models

Asian Spine J. 2024 May 20. doi: 10.31616/asj.2023.0404. Online ahead of print.

Abstract

Study design: Feasibility study.

Purpose: A phantom model was used to evaluate the accuracy of a novel augmented reality (AR) system for cervical screw placement.

Overview of literature: The use of navigation systems is becoming increasingly common in spine procedures. However, numerous factors limit the feasibility of regular and widespread use of navigation tools during spine surgery. AR is a new technology that has already demonstrated utility as a navigation tool during spine surgery. However, advancements in AR technology are needed to increase its adoption by the medical community.

Methods: AR technology that uses a fiducial-less registration system was tested in a preclinical cervical spine phantom model study for accuracy during spinal screw placement. A three-dimensional reconstruction of the spine along with trajectory lines was superimposed onto the phantom model using an AR headset. Participants used the AR system to guide screw placement, and post-instrumentation scans were compared for accuracy assessment.

Results: Twelve cervical screws were placed under AR guidance. All screws were placed in an acceptable anatomic position. The average distance error for the insertion point was 2.73±0.55 mm, whereas that for the endpoint was 2.71±0.69 mm. The average trajectory angle error for all insertions was 2.69°±0.59°.

Conclusions: This feasibility study describes a novel registration approach that superimposes spinal anatomy and trajectories onto the surgeon's real-world view of the spine. These results demonstrate reasonable accuracy in the preclinical model. The results of this study demonstrate that this technology can assist with accurate screw placement. Further investigation using cadaveric and clinical models is warranted.

Keywords: Augmented reality; Cervical; Mixed reality; Screw fixation.