Design of a handheld and portable fluorescence imaging system for quantitative detection of pregnancy-specific biomarkers in cattle

Anal Bioanal Chem. 2024 May 14. doi: 10.1007/s00216-024-05333-6. Online ahead of print.

Abstract

Reproductive management significantly impacts dairy farm productivity, necessitating accurate timely pregnancy detection in cattle. This paper presents a novel handheld and portable fluorescence imaging system designed for quantitative assessment of pregnancy-specific biomarkers, addressing the limitations of current detection methods. The objective was to develop a cost-effective, at-farm solution for detecting pregnancy-specific protein B (PSPB) in bovine plasma samples. The system integrates an imaging module and a custom software application, enabling image capture, data processing, and PSPB concentration determination. Calibration utilizing known PSPB concentrations achieved a 0.6 ng/mL limit of detection. Validation encompassed a comparison with a standard ELISA method using 100 bovine plasma samples; minimal bias and good agreement were observed within the linear range of the calibration curve for both methods. The system offers portability, user-friendliness, and potential for multiplex detection, promising real-time, at-farm reproductive management. This study demonstrates the successful development and validation of a portable fluorescence imaging system, offering an efficient and accurate approach to detecting pregnancy-specific biomarkers in cattle. Its implications extend to improving dairy farm productivity by enabling timely and reliable reproductive management practices.

Keywords: Bovine plasma; Fluorescence imaging; Lateral flow assay; Reproductive management.