Staged implantation of active transcutaneous bone conduction hearing devices (BCHD) after explantation of older generation bone anchored hearing aids (BAHAs): Surgical outcomes and approach to management

Am J Otolaryngol. 2024 Apr 28;45(4):104328. doi: 10.1016/j.amjoto.2024.104328. Online ahead of print.

Abstract

Purpose: To evaluate outcomes following explantation of percutaneous or transcutaneous bone conduction implants (pBCIs or tBCIs) and subsequent implantation of transcutaneous active bone conduction hearing devices (BCHDs); to provide guidance regarding staging of surgery and adjunctive procedures.

Materials and methods: Retrospective chart review of eight adult subjects (ten ears) with pBCIs or tBCIs who underwent explantation of their device and subsequent implantation with a BCHD [MED-EL BONEBRIDGE™ (n = 7, 70 %) or Cochlear™ Osia® (n = 3, 30 %)].

Results: Reasons for pBCI or tBCI explantation were pain (60 %, 6/10), infection (60 %, 6/10), skin overgrowth (50 %, 5/10), and inability to obtain new processors (20 %, 2/10). Median time between pBCI or tBCI removal and BCHD staged implant was 4.7 (IQR 2.2-8.1) months. Two subjects developed complications following BCHD implantation. One had a persistent wound overlying the osseointegrated screw after removal of the pBCI abutment, requiring removal and temporalis rotational flap. Staged Osia® implantation was performed, but ultimately wound dehiscence developed over the device. The second subject experienced an infection after BONEBRIDGE™ implantation (32 days after pBCI explant), necessitating washout and treatment with intravenous antibiotics. There was subsequent device failure.

Conclusion: The transition from a pBCI or tBCI to a novel transcutaneous device is nuanced. Staged pBCI or tBCI explantation and novel BCHD implantation with sufficient time for wound healing is vital. Adjunctive procedures to augment soft tissue in cases of prior attenuation may be required to avoid complications with larger internal devices.