Comparison of a portable, pneumotach flow-sensor-based spirometer (Spirofy™) with the vitalograph alpha Touch™ spirometer in evaluating lung function in healthy individuals, asthmatics, and COPD patients-a randomized, crossover study

BMC Pulm Med. 2024 May 10;24(1):230. doi: 10.1186/s12890-024-02972-4.

Abstract

Background: Spirofy™ is India's first portable, pneumotach flow-sensor-based digital spirometer developed to diagnose asthma and chronic obstructive pulmonary disease (COPD). In this study, we compared the performance of the Spirofy™ device with that of the Vitalograph Alpha Touch™ spirometer in measuring the lung capacities of healthy individuals, asthmatics, and COPD patients. We also assessed the inter-device variability between two Spirofy™ devices.

Methods: In a randomized, three-way crossover, open-label study, we measured the differences in forced expiratory volume in the first second (FEV1) and forced vital capacity (FVC) between the Spirofy™ and Vitalograph Alpha Touch™ spirometers. A proportion of the FEV1/FVC ratio distribution of < 0.7 was used to compare the diagnostic accuracies of the Spirofy™ with Vitalograph™ Alpha Touch™ spirometers.

Results: Ninety subjects participated in this study. The mean ± SD FVC values obtained from the Spirofy™ 1, Spirofy™ 2, and Vitalograph Alpha Touch™ devices were 2.60 ± 1.05 L, 2.64 ± 1.04 L, and 2.67 ± 1.04 L, respectively. The mean ± SD FEV1 values obtained from the Spirofy™ 1, Spirofy™ 2, and Vitalograph Alpha Touch™ devices were 1.87 ± 0.92 (L), 1.88 ± 0.92 (L), and 1.93 ± 0.93 (L), respectively. A significant positive correlation was found between the FVC and FEV1 values recorded by Vitalograph Alpha Touch™, Spirofy™ 1, and Spirofy™ 2. As compared to Vitalograph Alpha Touch™, the Spirofy™ device showed good sensitivity (97%), specificity (90%), and overall accuracy (93.3%) at an FEV1/FVC ratio < 0.7. No inter-device variability was observed between the two Spirofy™ devices.

Conclusion: Spirofy™ is a portable and easy-to-use device and is as accurate as the standard Vitalograph Alpha Touch™ spirometer for the diagnosis of COPD and asthma.

Trial registration: CTRI/2021/09/036492 (Clinical Trials Registry - India).

Keywords: Asthma; Chronic obstructive pulmonary disease; Forced expiratory volume; Forced vital capacity; Spirometer.

Publication types

  • Randomized Controlled Trial
  • Comparative Study

MeSH terms

  • Adult
  • Aged
  • Asthma* / diagnosis
  • Asthma* / physiopathology
  • Cross-Over Studies*
  • Female
  • Forced Expiratory Volume
  • Humans
  • India
  • Male
  • Middle Aged
  • Pulmonary Disease, Chronic Obstructive* / diagnosis
  • Pulmonary Disease, Chronic Obstructive* / physiopathology
  • Spirometry* / instrumentation
  • Vital Capacity