An Assessment of the Imaging Performance of Hand-Held Ultrasound Scanners Using the Edinburgh Pipe Phantom

Ultrasound Med Biol. 2024 May 4:S0301-5629(24)00174-1. doi: 10.1016/j.ultrasmedbio.2024.04.003. Online ahead of print.

Abstract

Objective: Although hand-held ultrasound devices (HHUSDs) are currently used for a diverse range of diagnostic and interventional applications the imaging performance of such scanners is rarely considered. The aim of this study was to assess the imaging performance of a wide-range of HHUSDs and compare their imaging performance to cart-based systems utilized for the same clinical applications.

Methods: The grayscale imaging performances of 19 HHUSDs from eight different manufacturers, manufactured between 2016 and 2021, were measured using a figure-of-merit known as the resolution integral. The imaging performance of the HHUSDs were compared to 142 cart-based ultrasound scanners.

Results: The HHUSD with the overall highest resolution integral (66) was a Butterfly (Burlington, MA, USA) wired phased array for small parts applications, followed by a Philips (Bothell, WA, USA) Lumify wired curvilinear transducer (57) for abdominal applications, a Butterfly wired phased array (56) for abdominal applications, a GE (Freiburg, Baden-Wurttemberg, Germany) VScan Air wireless linear array (56) for small parts applications, and a Healcerion (Seoul, Korea) Sonon 300L wireless linear array (56) for small parts applications. A GE VScan Extend wired phased array had the highest resolution integral (44) for cardiac applications.

Conclusions: The Butterfly phased array had the highest resolution integral of all the 19 HHUSDs, although this value is still less than the majority of cart-based cardiac and abdominal ultrasound scanners manufactured from 2010 to 2017. Clinical users of HHUSDs should be mindful of the limitations in imaging performance of hand-held ultrasound devices.

Keywords: Edinburgh Pipe Phantom; Handheld; Performance; Ultrasound device; Wired; Wireless.