Determination of optimum levels of binding antibody units (BAU) of new quantitative chemiluminescent immuno-assay (CLIA) in COVID-19 vaccinated volunteer blood donors

Transfus Apher Sci. 2024 Apr 24;63(3):103937. doi: 10.1016/j.transci.2024.103937. Online ahead of print.

Abstract

Background: For assessment of COVID-19 vaccine efficacy, neutralization activity of anti-SARS-CoV-2 antibody is measured. This study was undertaken to determine optimum levels of binding antibody units (BAU/ml) in new quantitative chemiluminescent assay (CLIA) that corresponded to neutralizing potential (30% inhibition) of sVNT assay.

Methods: Ninety-one blood samples were analyzed by CLIA and sVNT assays. Test samples (n = 75) were collected from blood donors post-2nd vaccination dose, while control samples (n = 16) were archived pre-COVID donor samples. Correlation between CLIA and sVNT was calculated and receiver operating characteristic (ROC) curve was drawn and analyzed.

Results: Results indicated excellent correlation between 57.5 BAU/ml on CLIA and 30%inhibition on sVNT assay. ROC curve analysis revealed that the area under the curve (AUC) was 0.971.

Discussion: The present study determined that 57.5 BAU/ml on CLIA corresponded to 30% inhibition on sVNT assay. Periodic quantitative analysis.

Keywords: IgG antibodies; Immunoassay; Neutralization; Quantitative CLIA; SVNT.