Optimizing Microfluidic Impedance Cytometry by Bypass Electrode Layout Design

Biosensors (Basel). 2024 Apr 19;14(4):204. doi: 10.3390/bios14040204.

Abstract

Microfluidic impedance cytometry (MIC) has emerged as a popular technique for single-cell analysis. Traditional MIC electrode designs consist of a pair of (or three) working electrodes, and their detection performance needs further improvements for microorganisms. In this study, we designed an 8-electrode MIC device in which the center pair was defined as the working electrode, and the connection status of bypass electrodes could be changed. This allowed us to compare the performance of layouts with no bypasses and those with floating or grounding electrodes by simulation and experiment. The results of detecting Φ 5 μm beads revealed that both the grounding and the floating electrode outperformed the no bypass electrode, and the grounding electrode demonstrated the best signal-to-noise ratio (SNR), coefficient of variation (CV), and detection sensitivity. Furthermore, the effects of different bypass grounding areas (numbers of grounding electrodes) were investigated. Finally, particles passing at high horizontal positions can be detected, and Φ 1 μm beads can be measured in a wide channel (150 μm) using a fully grounding electrode, with the sensitivity of bead volume detection reaching 0.00097%. This provides a general MIC electrode optimization technology for detecting smaller particles, even macromolecular proteins, viruses, and exosomes in the future.

Keywords: floating electrode; grounding electrode; microfluidic impedance cytometry; no bypass electrode; sensing sensitivity.

MeSH terms

  • Biosensing Techniques
  • Electric Impedance*
  • Electrodes*
  • Equipment Design
  • Flow Cytometry
  • Microfluidic Analytical Techniques
  • Microfluidics
  • Signal-To-Noise Ratio

Grants and funding

This research received no external funding.