Targeted CQA analytical control strategy for commercial antibody products: Replacing ion-exchange chromatography methods for charge heterogeneity with multi-attribute monitoring

MAbs. 2024 Jan-Dec;16(1):2341641. doi: 10.1080/19420862.2024.2341641. Epub 2024 Apr 23.

Abstract

Peptide mapping with mass spectrometry (MS) is an important tool for protein characterization in the biopharmaceutical industry. Historically, peptide mapping monitors post-translational modifications (PTMs) of protein products and process intermediates during development. Multi-attribute monitoring (MAM) methods have been used previously in commercial release and stability testing panels to ensure control of selected critical quality attributes (CQAs). Our goal is to use MAM methods as part of an overall analytical testing strategy specifically focused on CQAs, while removing or replacing historical separation methods that do not effectively distinguish CQAs from non-CQAs due to co-elution. For example, in this study, we developed a strategy to replace a profile-based ion-exchange chromatography (IEC) method using a MAM method in combination with traditional purity methods to ensure control of charge variant CQAs for a commercial antibody (mAb) drug product (DP). To support this change in commercial testing strategy, the charge variant CQAs were identified and characterized during development by high-resolution LC-MS and LC-MS/MS. The charge variant CQAs included PTMs, high molecular weight species, and low molecular weight species. Thus, removal of the IEC method from the DP specification was achieved using a validated LC-MS MAM method on a QDa system to directly measure the charge variant PTM CQAs in combination with size exclusion chromatography (SE-HPLC) and capillary electrophoresis (CE-SDS) to measure the non-PTM charge variant CQAs. Bridging data between the MAM, IEC, and SE-HPLC methods were included in the commercial marketing application to justify removing IEC from the DP specification. We have also used this MAM method as a test for identity to reduce the number of QC assays. This strategy has received approvals from several health authorities.

Keywords: Assay replacement; CQA monitoring; MAM; QDa; control strategy; mass spectrometry; monoclonal antibody; multi-attribute method; multi-attribute monitoring; peptide mapping.

MeSH terms

  • Antibodies, Monoclonal* / chemistry
  • Chromatography, Ion Exchange / methods
  • Humans
  • Peptide Mapping* / methods
  • Protein Processing, Post-Translational
  • Quality Control
  • Tandem Mass Spectrometry / methods

Substances

  • Antibodies, Monoclonal

Grants and funding

The author(s) reported there is no funding associated with the work featured in this article.