Detunable wireless resonator arrays for TMJ MRI: A comparative study

Magn Reson Imaging. 2024 Apr 14:111:84-89. doi: 10.1016/j.mri.2024.04.016. Online ahead of print.

Abstract

Temporomandibular Joint Magnetic Resonance Imaging (TMJ MRI) is crucial for diagnosing temporomandibular disorders (TMDs). This study advances the use of inductively coupled wireless coils to enhance imaging quality in TMJ MRI. After investigating multiple wireless resonator configurations, including a 1-loop design with a loop diameter of 9 cm, a 2-loop design with each loop having a diameter of 7 cm, and a 3-loop design with each loop having a diameter of 5 cm, our findings indicate that the 3-loop configuration achieves the optimal signal-to-noise ratio (SNR), surpassing other wireless arrays. Bilateral deployment of wireless coils further amplifies SNR, enabling superior visualization of TMJ structures, particularly with the 3-loop design. This cost-effective and comfortable solution, featuring a detunable design, eliminates the need for system parameter adjustments. The study indicates broad adaptability across MRI platforms, enhancing TMJ imaging for routine clinical diagnostics of TMDs.

Keywords: Array; Detune; MRI; RF coil; TMJ; Wireless coil.