Abstract

Magnetic Scanning Microscopy (MSM) emerged with the aim of allowing the visualization of magnetic fields of a sample or material through scanning and proved particularly useful for geology, biomedicine, characterization of magnetic materials, and in the steel industry. In this regard, the reading system of an MSM was modified using a μ-metal magnetic shielding structure to analyze remanent fields. The MSM was adapted to perform readings using two different types of sensors. The sensitive area of the sensors was evaluated, and the HQ-0811 (AKM-Asahi KaseiTM Microdevices) and STJ-010 (Micro MagneticsTM) sensors were chosen, with the HQ-0811 standardized on Printed Circuit Boards (PCBs) to facilitate handling and increase the system's robustness. In the shielded chamber, two piezoelectric ANC-150 stepper motors (Attocube Systems) were used, arranged planarly, to allow the movement of the analyzed samples under the mounted sensors. To acquire data from the sensors, the Precision Current Source Model 6220 and the Nanovoltmeter Model 2182A (both from Keithley) were used, along with Keithley's Delta-Mode integrated system. To analyze the system's effectiveness, three distinct samples were analyzed for calibration, and a MATLAB program was written to analyze the images and extract the material's magnetization. Additionally, a rock sample from the Parnaíba Basin was mapped to demonstrate the system's capabilities.