The dilution evaluation as a corrective measure for interference in the white blood cell scattergram in Beckman Coulter DxH 900

J Clin Lab Anal. 2024 Apr 9:e25007. doi: 10.1002/jcla.25007. Online ahead of print.

Abstract

Background: The Beckman Coulter DxH 900 is a haematological analyser capable of counting and sizing blood cells, and obtaining a complete blood cell count (CBC). This analyses different parameters of red blood cells (RBC), platelets and white blood cells/leukocytes. Some automated CBC counters present limitations due to specimen characteristics, abnormal cells or both factors. In the presence of abnormalities, the DxH 900 has a flagging system, warning the laboratory technician that something needs to be verified. In the present work, we evaluated samples from oncologic patients, presenting a population erroneously perceived as being lymphocytes. The most common explanations for this situation are RBC resistant to lysis or serum hyperbilirubinaemia.

Methods: In an attempt to solve and understand what the cause of this problem might be, we diluted our samples (1:3) and analysed the serum total bilirubin. To identify cells' abnormalities, the samples were also analysed by manual DLC counts. During the study, we also checked the different flags presented by the equipment.

Results: The results evidenced that the major interference was due to RBC lysis resistance, corresponding to 94.7% of the cases, while hyperbilirubinaemia was only present in 73.4%. Besides, we determined that some samples with normal bilirubin levels also presented interference, suggesting that hyperbilirubinaemia was not the main cause of the error. The most recurrent flag observed was "High event rate".

Conclusion: The dilution solved all of the observed interferences. The results between diluted and manual counts showed a strong correlation, leading us to introduce dilution in our laboratory routine.

Keywords: dilution; high event rate; hyperbilirubinaemia; red blood cell lysis resistance; white blood cell scattergram.