Improvement of point of care testing device for accurate whole blood glucose measurement in early neonates

Ann Clin Biochem. 2024 Apr 30:45632241249034. doi: 10.1177/00045632241249034. Online ahead of print.

Abstract

Background: It is important that blood glucose concentrations be accurately and conveniently measured in infants. However, especially in the early neonatal period, point-of-care testing devices used for adults may not accurately measure blood glucose concentrations in neonates.

Methods: In Study 1, the accuracy of neonatal whole-blood glucose measurements was evaluated for the existing glucose analyser Glutest Mint® (hereinafter MINT1; Sanwa Kagaku Kenkyusho, Nagoya, Japan) by comparing the data with reference blood glucose concentrations. In Study 2, we used MINT2, which was modified based on the findings from Study 1, to measure whole-blood glucose concentrations in newborns, and the accuracy of the measurements was compared with that of MINT1.

Results: Blood glucose concentrations were measured in 100 infants each in Study 1 and 2. In Study 1, the whole-blood glucose concentrations measured using MINT1 were found to be significantly lower than the reference blood glucose concentrations in early neonates. The results of Study 1 suggested that characteristics of erythrocyte membranes in early neonates affected the measurements. Therefore, we conducted Study 2 using MINT2, which was modified to be less susceptible. MINT2 was found to accurately measure whole-blood glucose concentrations in the early neonatal period.

Conclusion: The study showed that the point-of-care testing device could be improved to allow for accurate whole-blood glucose measurements in the early neonatal period.

Keywords: Point-of-care testing device; accuracy; neonates; whole-blood glucose concentration.