[High Purity Solvents for qNMR Measurement]

Yakugaku Zasshi. 2024;144(4):367-371. doi: 10.1248/yakushi.23-00151-4.
[Article in Japanese]

Abstract

Quantitative NMR (qNMR) has been adopted by documentary standards, including the Japanese Pharmacopoeia (JP), United States Pharmacopoeia (USP), and International Organization for Standardization (ISO), owing to its reliability and efficiency. Note that qNMR can be used for quantifying target components using the signal integration ratio of an analyte to a reference. In qNMR, a modern NMR instrument with high resolution and sensitivity is used to record reliable spectra. This instrument can detect small signals from impurities in a solvent, which may result in inaccurate signal integration in the spectrum. In this study, we investigated the influence of solvent quality on qNMR accuracy focusing on organic impurities, water content, and deuteration ratio. If signals from organic impurities and signals from the analyte overlap, the duplication of signal integration will directly affect the qNMR analytical result. To examine overlapping, we performed blank solvent tests. Additionally, a high water content and low deuteration ratio affect the detection sensitivity, thus reducing the signal-to-noise (S/N) ratio of the target. Thus, these factors must be considered to obtain accurate qNMR results.

Keywords: deuterated solvent; deuteration degree; quantitative NMR.

Publication types

  • English Abstract

MeSH terms

  • Magnetic Resonance Spectroscopy / methods
  • Reference Standards
  • Reproducibility of Results
  • Solvents
  • Water*

Substances

  • Solvents
  • Water