Quantitative analysis of iodine and bromine in soil using an energy-dispersive X-ray fluorescence spectrometer with three-dimensional polarized optics

Anal Sci. 2024 Mar 26. doi: 10.1007/s44211-024-00541-7. Online ahead of print.

Abstract

A nondestructive and simple method for the determination of bromine and iodine in soil was investigated using an energy-dispersive X-ray fluorescence spectrometer equipped with three-dimensional polarized optics. Using a gadolinium X-ray tube and Mo and Al2O3 as optimal secondary target materials, K-lines were detected for bromine and iodine. The minimum detection limits for bromine and iodine were calculated using JSAC0411, a certified reference material of soil for metal composition analysis, and were 0.77 mg/kg for bromine and 2.3 mg/kg for iodine at a measurement time of 600 s. The results of the determination of bromine and iodine in soil samples by the standard addition method were 256 ± 8 mg/kg for iodine and 67.9 ± 1.3 mg/kg for bromine with JSAC0411, which were in close agreement with the results measured by inductively coupled plasma mass spectrometry (ICP-MS) after alkaline extraction with tetramethylammonium hydroxide (TMAH) solution. The method developed in this study is an excellent technique for direct analysis of soil by X-ray fluorescence analysis without any pretreatment such as alkaline extraction. It is expected to be a practical analytical method for elucidating the dynamics of bromine and iodine in agricultural land and soil.

Keywords: Bromine; ED-XRF; Iodine; Soil; Standard addition method.