Three-Day Monitoring of Adhesive Single-Lead Electrocardiogram Patch for Premature Ventricular Complex: Prospective Study for Diagnosis Validation and Evaluation of Burden Fluctuation

J Med Internet Res. 2024 Mar 21:26:e46098. doi: 10.2196/46098.

Abstract

Background: Wearable electrocardiogram (ECG) monitoring devices are used worldwide. However, data on the diagnostic yield of an adhesive single-lead ECG patch (SEP) to detect premature ventricular complex (PVC) and the optimal duration of wearing an SEP for PVC burden assessment are limited.

Objective: We aimed to validate the diagnostic yield of an SEP (mobiCARE MC-100, Seers Technology) for PVC detection and evaluate the PVC burden variation recorded by the SEP over a 3-day monitoring period.

Methods: This is a prospective study of patients with documented PVC on a 12-lead ECG. Patients underwent simultaneous ECG monitoring with the 24-hour Holter monitor and SEP on the first day. On the subsequent second and third days, ECG monitoring was continued using only SEP, and a 3-day extended monitoring was completed. The diagnostic yield of SEP for PVC detection was evaluated by comparison with the results obtained on the first day of Holter monitoring. The PVC burden monitored by SEP for 3 days was used to assess the daily and 6-hour PVC burden variations. The number of patients additionally identified to reach PVC thresholds of 10%, 15%, and 20% during the 3-day extended monitoring by SEP and the clinical factors associated with the higher PVC burden variations were explored.

Results: The recruited data of 134 monitored patients (mean age, 54.6 years; males, 45/134, 33.6%) were analyzed. The median daily PVC burden of these patients was 2.4% (IQR 0.2%-10.9%), as measured by the Holter monitor, and 3.3% (IQR 0.3%-11.7%), as measured in the 3-day monitoring by SEP. The daily PVC burden detected on the first day of SEP was in agreement with that of the Holter monitor: the mean difference was -0.07%, with 95% limits of agreement of -1.44% to 1.30%. A higher PVC burden on the first day was correlated with a higher daily (R2=0.34) and 6-hour burden variation (R2=0.48). Three-day monitoring by SEP identified 29% (12/42), 18% (10/56), and 7% (4/60) more patients reaching 10%, 15%, and 20% of daily PVC burden, respectively. Younger age was additionally associated with the identification of clinically significant PVC burden during the extended monitoring period (P=.02).

Conclusions: We found that the mobiCARE MC-100 SEP accurately detects PVC with comparable diagnostic yield to the 24-hour Holter monitor. Performing 3-day PVC monitoring with SEP, especially among younger patients, may offer a pragmatic alternative for identifying more individuals exceeding the clinically significant PVC burden threshold.

Keywords: ECG; EKG; cardiology; cardiovascular; electrocardiogram; extended monitor; heart; premature ventricular complex; single-lead electrocardiogram; wearable; wearable device.

MeSH terms

  • Electrocardiography
  • Electrocardiography, Ambulatory
  • Humans
  • Male
  • Middle Aged
  • Prospective Studies
  • Technology
  • Ventricular Premature Complexes* / diagnosis