Redo-TAVR Feasibility After SAPIEN 3 Stratified by Implant Depth and Commissural Alignment: A CT Simulation Study

Circ Cardiovasc Interv. 2024 Mar;17(3):e013766. doi: 10.1161/CIRCINTERVENTIONS.123.013766. Epub 2024 Mar 19.

Abstract

Background: Redo-transcatheter aortic valve replacement (TAVR) can pin the index transcatheter heart valve leaflets open leading to sinus sequestration and restricting coronary access. The impact of initial implant depth and commissural alignment on redo-TAVR feasibility is unclear. We sought to determine the feasibility of redo-TAVR and coronary access after SAPIEN 3 (S3) TAVR stratified by implant depth and commissural alignment.

Methods: Consecutive patients with native valve aortic stenosis were evaluated using multidetector computed tomography. S3 TAVR simulations were done at 3 implant depths, sizing per manufacturer recommendation and assuming nominal expansion in all cases. Redo-TAVR was deemed unfeasible based on valve-to-sinotubular junction distance and valve-to-sinus height <2 mm, while the neoskirt plane of the S3 transcatheter heart valve estimated coronary access feasibility.

Results: Overall, 1900 patients (mean age, 80.2±8 years; STS-PROM [Society of Thoracic Surgeons Predicted Risk of Operative Mortality], 3.4%) were included. Redo-TAVR feasibility reduced significantly at shallower initial S3 implant depths (2.3% at 80:20 versus 27.5% at 100:0, P<0.001). Larger S3 sizes reduced redo-TAVR feasibility, but only in patients with a 100:0 implant (P<0.001). Commissural alignment would render redo-TAVR feasible in all patients, assuming the utilization of leaflet modification techniques to reduce the neoskirt height. Coronary access following TAV-in-TAV was affected by both index S3 implant depth and size.

Conclusions: This study highlights the critical impact of implant depth, commissural alignment, and transcatheter heart valve size in predicting redo-TAVR feasibility. These findings highlight the necessity for individualized preprocedural planning, considering both immediate results and long-term prospects for reintervention as TAVR is increasingly utilized in younger patients with aortic stenosis.

Keywords: aortic stenosis; balloon-expandable valve; lifetime management; transcatheter aortic valve replacement.

MeSH terms

  • Aged
  • Aged, 80 and over
  • Aortic Valve / diagnostic imaging
  • Aortic Valve / surgery
  • Aortic Valve Stenosis* / diagnostic imaging
  • Aortic Valve Stenosis* / surgery
  • Feasibility Studies
  • Heart Valve Prosthesis*
  • Humans
  • Prosthesis Design
  • Risk Factors
  • Tomography, X-Ray Computed
  • Transcatheter Aortic Valve Replacement* / adverse effects
  • Transcatheter Aortic Valve Replacement* / methods
  • Treatment Outcome