Application of high-sensitivity radon monitor on walk-in type radon calibration chamber at KRISS

Appl Radiat Isot. 2024 Jun:208:111285. doi: 10.1016/j.apradiso.2024.111285. Epub 2024 Mar 11.

Abstract

This paper introduces the KRISS-Rn4, a high-sensitivity radon monitor with four detection cells, installed within a walk-in type radon calibration chamber at KRISS. The KRISS-Rn4 exhibits enhanced energy resolution through channel-by-channel signal processing and data acquisition. Results reveal that it achieves lower statistical fluctuations and faster response times in monitoring test atmospheres compared to commercial devices.

Keywords: Alpha spectroscopy; KRISS-Rn4; RAD7; Radon; Radon monitor; Radon monitor calibration; Walk-in radon chamber.