Four-Point Impedance: A Potential Biomarker for Residual Hearing After Cochlear Implantation

Otol Neurotol. 2024 Apr 1;45(4):e315-e321. doi: 10.1097/MAO.0000000000004153.

Abstract

Introduction: Preservation of residual hearing after cochlear implantation allows for electroacoustic stimulation, which leads to better music appreciation, noise localization, and speech comprehension in noisy environments. Real-time intraoperative electrocochleography (rt-ECochG) monitoring has shown promise in improving residual hearing rates. Four-point impedance (4PI) is being explored as a potential biomarker in cochlear implantation that has been associated with fibrotic tissue response, hearing loss, and dizziness. In this study, we explore whether monitoring both rt-ECochG intraoperatively and postoperative 4PI improves predictions of the preservation of residual hearing.

Methods: This was a prospective cohort study. Adults with residual acoustic hearing underwent cochlear implantation with intraoperative intracochlear electrocochleography (ECochG) monitoring. The surgeon responded to a drop in ECochG signal amplitude of greater than 30% by a standardized manipulation of the electrode with the aim of restoring the ECochG. At the end of the procedure, the ECochG signal was categorized as being maintained or having dropped more than 30%. 4PI was measured on 1 day, 1 week, and 1 and 3 months after cochlear implantation. Residual hearing was measured by routine pure-tone audiogram at 3 months postoperatively. The ECochG category and 4PI impedance values were entered as factors in a multiple linear regression predicting the protection of residual hearing.

Results: Twenty-six patients were recruited. Rt-ECochG significantly predicted residual hearing at 3 months (t test; mean difference, 37.7%; p = 0.002). Inclusion of both 1-day or 3-month 4PI in a multiple linear regression with rt-ECochG markedly improved upon correlations with residual hearing compared with the rt-ECochG-only model (rt-ECochG and 1-d 4PI model, R2 = 0.67; rt-ECochG and 3-mo 4PI model, R2 = 0.72; rt-ECochG-only model, R2 = 0.33).

Conclusions: Both rt-ECochG and 4PI predict preservation of residual hearing after cochlear implantation. These findings suggest that the biological response of the cochlea to implantation, as reflected in 4PI, is an important determinant of residual hearing, independent of the acute effects on hearing during implant surgery seen with rt-ECochG. We speculate that 4PI relates to inflammation 1 day after implantation and fibrosis at 3 months.

MeSH terms

  • Adult
  • Audiometry, Evoked Response / methods
  • Biomarkers
  • Cochlea / surgery
  • Cochlear Implantation* / methods
  • Cochlear Implants*
  • Electric Impedance
  • Hearing
  • Humans
  • Prospective Studies

Substances

  • Biomarkers