Toward a New Primary Standardization of Radionuclide Massic Activity Using Microcalorimetry and Quantitative Milligram-Scale Samples

J Res Natl Inst Stand Technol. 2022 Feb 24:126:126048. doi: 10.6028/jres.126.048. eCollection 2021.

Abstract

We present a new paradigm for the primary standardization of radionuclide activity per mass of solution (Bq/g). Two key enabling capabilities are 4π decay-energy spectrometry using chip-scale sub-Kelvin microcalorimeters and direct realization of mass by gravimetric inkjet dispensing using an electrostatic force balance. In contrast to traditional traceability, which typically relies on chemical separation of single-radionuclide samples, 4π integral counting, and additional spectrometry methods to verify purity, the system described here has both 4π counting efficiency and spectroscopic resolution sufficient to identify multiple radionuclides in the same sample at once. This enables primary standardization of activity concentrations of mixed-radionuclide samples. A major benefit of this capability, beyond metrology, is in assay of environmental and forensics samples, for which the quantification of multiplenuclide samples can be achieved where presently inhibited by interferences. This can be achieved without the need for chemical separations or efficiency tracers, thereby vastly reducing time, radioactive waste, and resulting measurement uncertainty.

Keywords: alpha; beta; cryogenic detectors; mass metrology; microcalorimeter; radioactivity; radionuclide metrology; transition edge sensor.